AEP transducers / CLSTM / Force transducers

Column force transducer suitable for the calibration of material testing machines.
High mechanical resistance.
Completely made of stainless steel.
Completely LASER welded.
Degree of protection IP65.
Long term stability.

Category:

Description

Force transducers for the measurement of static and dynamic loads in COMPRESSION

Nominal Load from 300 kN to 5 MN
Class (ISO 376) 1 and 0.5
Linearity and Hysteresis 0,05%
Output: 2mV/V

Dimensions[mm]

Load Æ A Æ B Æ C D E F Æ G H L M Æ N P R T Z
30t, 60t, 100t 102 94 76 15 15 70 102 165 65 M8 79 12 200 / 16
200t, 300t 129 122 106 20 14 90 129 215 80 M12 109 17 350 / 18
500t 168 160 140 35 35 130 168 250 100 M16 143 21 500 M8 20

Technical Data

Accuracy Class C1 C2
Legal Divisions 1000 2000
Nominal Load (Emax) 30, 60, 100, 200, 300, 500 t
Min. Verification Interval (Vmin) Emax / 10000
Combined Error ±0.045% ±0.023%
Non-Repeatability ±0.015% ±0.010%
Zero Return over 30 min. ±0.036% ±0.026%
Creep at nominal load:
a) over 30 min. ±0.035% ±0.028%
b) over 20 and 30 min. ±0.010% ±0.008%
Temperature Effect (10 °C)
a) on zero
b) on sensitivity
±0.030%
±0.030%
±0.024%
±0.017%
Nominal Sensitivity Calibration Tolerance 2 mV/V ±0.1%
Nominal Load 30, 60, 100 t
Input Resistance 700 ± 2 Ω
Output Resistance 700 ± 2 Ω
Nominal Load 200, 300, 500 t
Input Resistance 350 ± 2 Ω
Output Resistance 350 ± 2 Ω
Insulation Resistance Zero Balance > 5 GW ±1% FS
Recommended Supply Voltage Nominal Supply Voltage 10 V, 1–15 V
Maximum Supply Voltage 18 V
Mechanical Limit values referred to nominal load:
a) minimum load 0%
b) service load 120%
c) max permissible load 150%
d) breaking load > 300%
e) max transverse load 50%
Reference Temperature +23 °C
Temperature Nominal Range -10 / +40 °C
Working Temperature -20 / +70 °C
Storage Temperature -20 / +80 °C
Nominal Load / Weight 30, 60, 100 t / ~7 kg 200, 300 t / ~16 kg 500 t / ~31 kg
Protection Class (EN 60529) Execution Material / Cable Length IP68, Acciaio Inox / Stainless Steel, 5 m (on request 10 m)

Acceleration of gravity g=9.80434 m/s2.

Electrical Connections
STANDARD OUTPUT: PVC 105°C shielded cable, Ø 5.2mm with 6 × Ø 0.25mm2 tinned conductors.


Shield connected to the body of the load cell.

Reviews

There are no reviews yet.

Be the first to review “AEP transducers / CLSTM / Force transducers”