Description
Force transducers for the measurement of static and dynamic loads in COMPRESSION and TENSION
Nominal Load from 500 N to 5 kN
Class (ISO 376) 00, 0.5 and 1
Linearity and Hysteresis 0,025%
Output: 2mV/V
On request


Dimensions[mm]

| CODE (Class 00) | CODE (Class 0.5) | CODE (Class 1) | LOAD |
| CTSTM63500N005 | CTSTM63500NI05 | CTSTM63500NI15 | 0.5 kN |
| CTSTM631KNI005 | CTSTM631KNI05 | CTSTM631KNI15 | 1 kN |
| CTSTM632KNI005 | CTSTM632KNI05 | CTSTM632KNI15 | 2 kN |
| CTSTM633KNI005 | CTSTM633KNI05 | CTSTM633KNI15 | 3 kN |
| CTSTM635KNI005 | CTSTM635KNI05 | CTSTM635KNI15 | 5 kN |
Technical Data
| Accuracy class: ISO 376 | 00 | 0.5 | 1 |
| NOMINAL LOAD | 0.5 – 1 – 2 – 3 – 5 kN | ||
| RELATIVE ERROR (at reading) a) repeatability 0°-120°-240° (b’) b) interpolation (fc) c) reversibility (u) d) zero (fo) |
≤0.050%(1) ≤0.020%(1) ≤0.070%(1) ≤0.010% F.S. |
≤0.090%(1) ≤0.040%(1) ≤0.100%(1) ≤0.025% F.S. |
≤0.155%(1) ≤0.080%(1) ≤0.200%(1) ≤0.025% F.S. |
| LINEARITY HYSTERESIS |
≤0.02% F.S. ≤0.025% F.S. |
||
| TEMPERATURE EFFECT (10°C) a) on zero b) on sensitivity |
≤0.025% F.S. ≤0.025% F.S. |
||
| NOMINAL SENSITIVITY SENSIVITY TOLERANCE |
2 mV/V ≤0.1% F.S. |
||
| INPUT RESISTANCE OUTPUT RESISTANCE INSULATION RESISTANCE ZERO BALANCE RECOMMENDED SUPPLY VOLTAGE NOMINAL SUPPLY VOLTAGE MAXIMUM SUPPLY VOLTAGE |
350 ± 2W 352 ± 2W >5 GW ≤1% F.S. 10 V 1-15 V 18 V |
||
| MECHANICAL LIMIT VALUES REFERRED TO NOMINAL LOAD: a) service load b) max permissible load c) breaking load d) max transverse load e) max permissible dynamic load DISPLACEMENT AT NOMINAL LOAD |
120% 150% >300% 50% 50% ~0.2 mm |
||
| REFERENCE TEMPERATURE TEMP. NOMINAL RANGE SERVICE TEMPERATURE STORAGE TEMPERATURE |
+23 °C -10 / +40 °C -10 / +70 °C -20 / +80 °C |
||
| WEIGHT PROTECTION CLASS (EN 60529) EXECUTION MATERIAL CABLE LENGTH |
~0.60 kg IP67 (0.5 kN IP65 silicon cover) Acciaio Inox / Stainless Steel 5 m |
||
(1) Percentage errors referred to reading, min. 1/ 10 of the nominal load
Classifications according ASTM E74 on request
Electrical Connections
STANDARD OUTPUT: PVC 105°C shielded cable, Ø 5.2mm with 4× Ø 0.35mm2 tinned conductors.

Shield connected to the body of the transducer.


Reviews