Description
BCL-A Single Point load cell
LOAD CELL TYPE: BCL-300GM ~ 4.5K-A
SPECIFICATIONS
PERFORMANCE
| Category | Item | Specification |
|---|---|---|
| Performance | Rated Capacity (R.C.) | 2.942 N (300 gf), 5.884 N (600 gf), 11.77 N (1.2 kgf), 14.71 N (1.5 kgf), 29.42 N (3 kgf), 44.13 N (4.5 kgf) |
| Safe Overload | 150 % R.C. | |
| Ultimate Overload | 200 % R.C. | |
| Rated Output (R.O.) | 1.6 mV/V ± 0.2 mV/V | |
| Non-linearity | 0.02 % R.O. | |
| Hysteresis | 0.02 % R.O. | |
| Repeatability | 0.02 % R.O. | |
| Creep | 0.02 % R.O. / 20 min | |
| Creep Recovery | 0.02 % R.O. / 20 min | |
| Electrical | Recommended Excitation | 10 V or less |
| Maximum Excitation | 15 V | |
| Zero Balance | ±0.04 mV/V | |
| Input Terminal Resistance | 420 Ω (+30 Ω / −20 Ω) | |
| Output Terminal Resistance | 350 Ω ± 5 Ω | |
| Insulation Resistance (Bridge–Body) | 2000 MΩ or more (DC 50 V) | |
| Temperature | Compensated Temperature Range | −10 °C to +50 °C |
| Safe Temperature Range | −10 °C to +50 °C | |
| Temperature Effect on Zero Balance | 0.04 % R.O. / 10 °C | |
| Temperature Effect on Output | 0.02 % LOAD / 10 °C | |
| Eccentric Load | Max. Size of Loading Plate | 200 × 200 mm (300 g ~ 600 g)300 × 300 mm (1.2 kg ~ 4.5 kg) |
| Center of Loading Plate | Loading plate must be adjusted to the center of the load cell | |
| Measuring Accuracy | Accuracy within 0.02 % R.O. with 1/2 rated load applied at:50 mm from center (300 g ~ 600 g)75 mm from center (1.2 kg ~ 4.5 kg) | |
| Others | Cable | Ø3.2 mm, 4-core shielded, 40 cm |
| IP Level | IP40 | |
| Material of Element | Aluminium Alloy | |
| Weight | Approx. 40 g |




Reviews