Description
Model 1140 Stainless Steel Single-Point Load Cell
FEATURES
• Capacities 15–150 kg
• Stainless steel construction
• Single-point 400 x 400 mm platform
• IP65 protection
• Available with UNC threads only
• Optional
❍ FM approval available
APPLICATIONS
• Harsh environment small platforms
• Harsh environment check weighing

| SPECIFICATIONS | ||
| PARAMETER | VALUE | UNIT |
| Rated capacity—R.C. (Emax) | 15, 20, 30, 50, 75, 100, 150 | kg |
| NTEP/OIML accuracy class | Non-Approved | |
| Maximum no. of intervals (n) | 3000 | |
| Rated output—R.O. | 2.0 | mV/V |
| Rated output tolerance | 0.2 | ±mV/V |
| Zero balance | 0.2 | ±mV/V |
| Zero return, 30 min. | 0.0170 | ±% of applied load |
| Total error (per OIML R60) | 0.0200 | ±% of rated output |
| Temperature effect on zero | 0.004 | ±% of rated output/°C |
| Temperature effect on output | 0.0010 | ±% of load/°C |
| Eccentric loading error | 0.0074 | ±% of rated load/cm |
| Temp. range, compensated | –10 to +40 | °C |
| Temp. range, safe | –20 to +70 | °C |
| Maximum safe central overload | 150 | % of R.C. |
| Ultimate central overload | 300 | % of R.C. |
| Excitation, recommended | 10 | VDC or VAC RMS |
| Excitation, maximum | 15 | VDC or VAC RMS |
| Input impedance | 385±15 | Ω |
| Output impedance | 350±3 | Ω |
| Insulation resistance | >2000 | MΩ |
| Cable length | 1.0 | m |
| Cable type | 6-wire, PVC, single loating screen | Standard |
| Construction | Stainless steel | |
| Environmental protection | IP65* | |
| Platform size (max) | 400 x 400 | mm |
| Recommended torque | Up to 30 kg: 7.0 50 kg and above: 10.0 | N*m |
* IP67 available on request
All speciications are subject to change without notice.