SHOWA SOKKI / DBJ Tension & Compression Load Cell

Coaxial beam type,Ultralow-Capacity,High Accuracy,General Purpose

Brand:

Description

DBJ Tension & Compression Load Cell

Features

  • Covering the minute wide capacity range 1 to 20N.
  • Using special designed 350Ω strain gages for the detection of load, thereby, an excellent linearity or thermal coefficient of 0.025%RO can be attained.
  • Attractive load charactetistics in maximum overload (200%RO) and in side load (40%).
  • Guaranteeing the load leaving off coefficient up to 5 cm maximum and within 0.005%/cm.
  • A specially designed mounting table (Optionally supplied) can be utilized as a measuring unit.

Appearance Dimensions

Model and Capacity / Dimension / etc. (unit : mm)

Model Capacity Deflection Natural frequency Weight
DBJ-1N 1 N 0.42mm 82 Hz 125 g
DBJ-2N 2 N 0.35mm 130 Hz 125 g
DBJ-5N 5 N 0.24mm 180 Hz 125 g
DBJ-10N 10 N 0.20mm 360 Hz 125 g
DBJ-20N 20 N 0.16mm 570 Hz 125 g

*Mass does not include cables.

 

Specifications

Safe Overload 200 %R.C.
Maximum permissible overload 500 %R.C.
Acceptable overload of stopper 40 %R.C.
Rated Output 2 mV/V±5%
Nonlinearity 0.025 %R.O.
Hysteresis 0.025 %R.O.
Load leaving off coefficient 0.005 %/cm
Excitation Voltage 7 V  (or less)
Safe Excitation Voltage 10 V
Input Rsistance 350 Ω
Output Rsistance 350 Ω
Compensated Temp.Range -10~45 ℃
Safe Temp.Range -20~60 ℃
Temp.Effect on Zero 0.004 %R.O./℃
Temp.Effect on Output 0.003 %/℃
Cable Φ4mm-4wire shielded cable, length : 2m(tip peeled off)
  • Acceptable overload of stopper: The acceptable overload coefficient of stopper represents the maximum permissible overload value exceeding which the load cell in question can still retain its status as a load cell. In other words, any influences may not be produced in that load cell even in case when internal mechanisms might be damaged within this overload range.
  • Maximum Side Load: The maximum side load coefficient represents the permissible load by which a load cell can endure at 90 degree angles with the loading axis of that load cell and it corresponds to 23.5 degrees if expressed by a slant angle against the loading point. Its sensitivity varies in proportion with “cosθ” but its varying accuracy is guaranteed.
  • Load Leaving Off Coefficient: The load leaving off coefficient indicates the amount of variations in the output taking place when the load center leaves off from the axial center of the load cell. The leaving off distance of Model DBJ is guaranteed to be 5 cm in the maximum.

Reviews

There are no reviews yet.

Be the first to review “SHOWA SOKKI / DBJ Tension & Compression Load Cell”