AEP transducers / CM-AMP / Amplified Force transducers

Transductor de fuerza compacto a bajo perfil.
Elevada resistencia mecánica.
Realizada por entero en acero inoxidable.
Completamente saldada con el LASER.
Protección IP67.

Category:

Description

Force transducers for the measurement of static and dynamic loads in COMPRESSION.

Nominal load from 50 kg to 5 t
Linearity and hysteresis 0.20%
COMPACT DIMENSIONS
Nominal sensitivity 2 mV/V

Loading head CTICCM42

 

Dimensions[mm]

NOMINAL LOAD CODE
50 kg CCM4250K3
100 kg CCM42100K3
250 kg CCM42250K3
500 kg CCM42500K3
1000 kg CCM421T3
2500 kg CCM422T53
3500 kg CCM423T53
5000 kg CCM425T3

Technical Data

NOMINAL LOAD 50 – 100 – 250 – 500 kg
1000 – 2500 – 3500 – 5000 kg
LINEARITY
HYSTERESIS
REPEATABILITY
≤ ± 0.20 % F.S.
≤ ± 0.20 % F.S.
≤ ± 0.05 % F.S.
NOMINAL SENSITIVITY
SENSIVITY TOLERANCE
4-20 mA (2 wires) or 0,5…10,5 V
< ±0.1%
NOMINAL POWER SUPPLY 10 V → 15-24 Vdc
4-20 mA → 18-24 Vdc
MAX. POWER SUPPLY
MAX. ABSORPTION (without load applied)
LOADING RESISTANCE:

a) tension
b) current
INSULATION RESISTANCE ZERO BALANCE
RESPONSE FREQUENCY
28Vdc
30mA
min. 3KW
from 0 to 470W
>2 GW
± 1%
~ 1 kHz
REFERENCE TEMPERATURE
TEMPERATURE NOMINAL
RANGE SERVICE TEMPERATURE

STORAGE TEMPERATURE
+23 °C
-10 / +40 °C
-20 / +70 °C
-20 / +80 °C
TEMPERATURE EFFECT (10 °C):
a) on zero
b) on sensitivity
≤ ± 0.030 % F.S.
≤ ± 0.025 % F.S.
MECHANICAL LIMIT VALUES:
a) service load
b) max permissible load
c) breaking load
d) max transverse load
e) max permissible dynamic load
DISPLACEMENT AT NOMINAL LOAD
120 % F.S.
150 % F.S.
>300 % F.S.
50 % F.S.
50 % F.S.
~ 0.2 mm
WEIGHT
PROTECTION CLASS (EN 60529)
EXECUTION MATERIAL

CABLE LENGTH
~ 0.20 kg
IP67
Acciaio inox / Stainless Steel
3 m

N calibrations on request.

Electrical Connections

PVC 70°C shielded cable, Ø 4.2mm with 4 tinned Ø 0.14mm2 conductors. Shield connected to the body of the transducer.