AEP transducers / MICROTC4-AMP / Amplified Force transducers

Force transducer suitable for material testing machines and test benches in general.
High mechanical resistance.
Completely made of stainless steel.
Completely LASER welded.
Degree of protection IP67.
Long term stability.

Category:

Description

Force transducers for the measurement of static loads in COMPRESSION and TENSION

Nominal Load from 1 kN to 10 kN
Linearity and Hysteresis 0,05%

SMALL SIZE

Standar output

 

microTC4 10V microTC4 + base

Dimensions[mm]

CODE LOAD
CTC4631KN 1 kN
CTC4632kN5 2,5 kN
CTC4635KN 5 kN
CTC46310KN 10 kN
CTC463100K 100 kg
CTC463250K 250 kg
CTC463500K 500 kg
CTC4631T 1 t

Technical Data

NOMINAL LOAD 1 – 2,5 – 5 – 10 kN
LINEARITY
HYSTERESIS
£ ±0.05% F.S.
£ ±0.05% F.S.
NOMINAL SENSITIVITY
SENSIVITY TOLERANCE
± 10 Volt (1)
£ ±0.1% F.S.
NOMINAL POWER SUPPLY
MAX. POWER SUPPLY
MAX. ABSORPTION
LOADING RESISTANCE
INSULATION RESISTANCE
ZERO BALANCE

RESPONSE FREQUENCY
18-24Vdc
28Vdc
30mA
min. 3KW
>2 GW
± 1%
~ 1 kHz
REFERENCE TEMPERATURE
TEMPERATURE NOMINAL
RANGE SERVICE TEMPERATURE

STORAGE TEMPERATURE
+23°C
-10 / +40 °C
-10 / +70 °C
-20 / +80 °C
TEMPERATURE EFFECT (10°C)
a) on zero
b) on sensitivity
£ ±0.028% F.S.
£ ±0.024% F.S.
MECHANICAL LIMIT VALUES REFERRED TO NOMINAL LOAD:
a) service load
b) max permissible load
c) breaking load
d) max transverse load
e) max permissible dynamic load
120%
150%
>300%
100%
75%
Displacement at nominal load EFFECT OF TRANSVERSE LOAD:
a) at 10% of nominal load
~0.06 mm
£ ±0.030% F.S.
WEIGHT 0.28 kg
PROTECTION CLASS (EN 60529)
EXECUTION MATERIAL
CONNECTOR OUTPUT
IP67
Acciaio Inox / Stainless Steel
M12 5 Poles Male connector
FIXING SCREWS:
a) diameter
b) resistance class
c) tightening torque
M5
12.9
8 Nm

(1) Tests and calibrations performed in COMPRESSION with the transducer mounted on a bearing support with correctly tightened clamping screws.
kg calibration on request.

Electrical connection 

 