Minebea intec / Load cell PR 6241, converter Connexx® and mounting kits PR 6041/PR 6043/PR 6044 For the precise weighing of process vessels

The PR 6241 series compression S-type load cells are specially designed for weighing small and medium process vessels and for high-precision dosing. The PR 6241 is produced entirely from stainless steel. A special design feature is the combination of ‘real’ shearing force principle and pendulum support principle for the utmost precision.

Tag:
Brand:

Description

Load cell PR 6241, converter Connexx® and mounting kits PR 6041/PR 6043/PR 6044 For the precise weighing of process vessels

Technical specifications

Load cell PR 6241

Parameter Description Abbr. D1 C3 C6 Unit
Accuracy class 0.04 0.015 0.008 %Emax
Minimum dead load Lowest limit of specified measuring range Emin 0 0 0 %Emax
Maximum capacity Highest limit of specified measuring range Emax See Ordering information table kg
Safe load limit Maximum load without irreversible damage Elim 150 150 150 %Emax
Destructive load Danger of mechanical destruction Ed > 300 > 300 > 300 %Emax
Minimum load cell verification interval (OIML)

For NTEP, see separate table

Minimum load cell verification interval, vmin = Emax/Y Y 5,000 14,000 20,000
Deadload output return Factor for deadload output return after load (DR=1/2*Emax/Z) Z 1,000 3,000 8,000
Rated output Relative output at maximum capacity Cn 2 2 2 mV/V
Tolerance on rated output Permissible deviation from rated output dc <0.25 <0.07 <0.07 %Cn
Zero output signal Load cell output signal under unloaded condition Smin 0±1.0 0±1.0 0±1.0 %Cn
Reproducibility Max. change in load cell output for repeated loading eR <0.01 <0.005 <0.005 %Cn
Creep Max. change of output signal at Emax for 30 min. dcr <0.03 <0.015 <0.008 %Cn
Non-linearity1) Max. deviation from best straight line through zero dLin <0.03 <0.01 <0.01 %Cn
Hysteresis1) Max. difference in LC output between loading and unloading dhy <0.04 <0.015 <0.008 %Cn
Temperature effect on Smin Max. change of Smin in BT TKSmin <0.028 <0.01 <0.007 %Cn/10 K
Temperature effect on C1) Max. change of C in BT TKC <0.03 <0.01 <0.005 %Cn/10 K
Input impedance Between supply terminals RLC 650 ± 6 650 ± 6 650 ± 6
Output impedance Between measuring terminals RO 610 ± 1 610 ± 0.5 610 ± 0.5
Insulation impedance Between measuring circuit and housing at 100 VDC RIS > 5,000 ×106 > 5,000 ×106 > 5,000 ×106
Insulation voltage Between circuit and housing (PR 6241/..E only) 500 500 500 V
Recommended supply voltage To hold the specified performance Bu 4 … 24 4 … 24 4 … 24 V
Max. supply voltage Permissible for continuous operation without damage Umax 28/(EX versions: 25) V
Nominal ambient temp. range To hold the specified performance BT -10 … +55 -10 … +55 -10 … +55 °C
Usable ambient temp. range Permissible for continuous operation without damage BTu -40 …+95 -40 …+95 -40 …+95 °C
Storage temperature range Without electrical and mechanical stress BTi -40 …+95 -40 …+95 -40 …+95 °C
Permissible eccentricity Permissible displacement from nominal load line Sex 5 5 5 mm
Vibration resistance Resistance against oscillations (IEC 68-2-6 Fc) 20 g, 100 h, 10 …150 Hz
Barometric pressure influence Influence of barometric pressure on output PKSmin <0.005 <0.0025 <0.0025 %Cn/kPa
Nominal deflection Max. elastic deformation under maximum capacity Snom < 0.3 for Emax < 5 t/< 0.5 for Emax = 5 t mm
Material (sensor) Stainless steel 1.4542
Protection class PR 6241: IP68*/IP69

Connexx®: IP65/IP68**

Cable Length: 5 m

Diameter: 5 mm

Cross section: 4 × 0.35 mm2 Material cable sheath: TPE

Bending radius ≥ 25 mm at fixed installation

≥ 75 mm at flexible installation

Connexx®
Nominal ambient temp. range To hold the specified performance BT -10 … +40 °C
Usable temperature range Permissible for continuous operation without damage BTu -30 … +60 °C
Storage temperature range Without electrical and mechanical stress BTi -30 … +70 °C

* The load cell can be submerged in water at a depth of 1.5 m for 10,000 hours.
** The module can be submerged in water at a depth of 1.5 m for 100 hours.
1) The data for Non-linearity, hysteresis and TKC are typical values. For OIML R60 or NTEP approved load cells the sum of these values is within the permissible cumulative error

NTEP classification PR 6241
NTEP NTEP minimum load cell verification interval, vmin
Maximum capacity with OIML class divisions nmax 100 kg 200 kg 300 kg 500 kg 1 t 2 t 3 t 5 t Unit
Class III Multiple D1 2,000 20 40 60 100 200 400 600 1,000 g
Class III Multiple C3 5,000 14.3 21 36 71 143 214 357 g
Class III Multiple C6 8,000 10 15 25 50 100 150 g
Class IIIL Multiple D1 5,000 6.7 13.3 20 33 67 133 200 333 g
Class IIIL Multiple C3 10,000 5 7.1 12 24 48 71 119 g
Class IIIL Multiple C6 10,000 3 5 8 17 33 50 g
Mounting kits PR 6041/PR 6043/PR 6044
Maximum capaci- ty of load cell

PR 6241

 

 

Mounting kit/ accessories

 

 

 

Material

 

 

Description/ comments

 

 

Mounting screws

Installation height (mm) Max. permissible horizontal force (kN)  

Horizontal dest- ructive force [KN]

 

Max. permissible lift-off force [kN]

Max. permissible vertical load with- out load cell
100 kg–

500 kg

 

PR 6041/04N

 

Elastomeric bearing

 

M12

 

53

 

/

 

/

 

/

 

/

 

100 kg–2 t

PR 6041/30N Electrogalvanised steel S235, passivated and sealed Plate mounting kit M8-5.8  

 

 

 

 

92

/ / / /
PR 6041/30S Stainless steel 1.4301

(AISI 304)

Plate mounting kit M8-A2-70 / / / /
PR 6043/30N Electrogalvanised steel S235, passivated and sealed MiniFLEXLOCK M8-8.8 6 20 / 2 t
PR 6043/30S  

 

Stainless steel 1.4301

(AISI 304)

MiniFLEXLOCK M8-A2-70 6 20 / 2 t
PR 6044/00S Plate mounting kit M8-A2-70 / / 10 2 t
PR 6044/10S MaxiFLEXLOCK M8-A2-70 6 20 10 2 t
PR 6041/51S Platform base / 44 / / / /
PR 6043/31S Load disc kit / / / / / /
 

3 t–5 t

PR 6041/40N Electrogalvanised steel S235, passivated and sealed Plate mounting kit M12-5.8  

 

 

 

 

130

/ / / /
PR 6041/40S Stainless steel 1.4301

(AISI 304)

Plate mounting kit M12-A2-70 / / / /
PR 6043/40N Electrogalvanised steel S235, passivated and sealed MiniFLEXLOCK M12-8.8 18 50 / 5 t
PR 6043/40S  

Stainless steel 1.4301

(AISI 304)

MiniFLEXLOCK M12-A2-70 18 50 / 5 t
PR 6044/01S Plate mounting kit M12-A2-70 / / 20 5 t
PR 6044/11S MaxiFLEXLOCK M12-A2-70 18 50 20 5 t

Converter Connexx®

Fitted with the converter Connexx®, the load cell PR 6241 offers a whole host of possibilities:

  •  The digital version guarantees fast signalling times for reliable dosing processes.
  •  Thanks to the use of field bus communication, the cable junction box is no longer needed.
  •  The standard interface CANopen ensures straightforward installation and calibration.
  •  CANopen allows for extra-long communication paths of up to 200 m.
  •  Weight values can also be generated individual- ly – per load cell – meaning that defective load cells can be identified more easily.

Technical diagrams

Compression S-type load cell PR 6241

Model A [mm] B [mm] C [mm] D [mm] E [mm]
PR 6241/12…52 60 65 50 23 M12
PR 6241/13…23 60 65 50 30 M12
PR 6241/33 90 95 64 30 M20 × 1.5
PR 6241/53 90 95 64 35 M20 × 1.5

Converter Connexx®

Mounting kits PR 6041/PR 6043/PR 6044

Elastomeric bearing PR 6041/04N

Plate mounting kit PR 6041/30N+S

Plate mounting kit PR 6041/40N+S

 

Platform base PR 6041/51

MiniFLEXLOCK PR 6043/30N+S

 

MiniFLEXLOCK PR 6043/40N+S

Plate mounting kit PR 6044/00S

Plate mounting kit PR 6044/01S

 

 

MaxiFLEXLOCK PR 6044/10S

MaxiFLEXLOCK PR 6044/11S

Ex approval
Scope of validity:
PR 6241 (100 kg…5 t)

Certificates for compression S-type load cell PR 6241
Zone Labelling Certificate number For
0 and 1 II 1G Ex ia IIC T6 Ga

Ex ia IIC T6 Ga

BVS 16 ATEX E 005

IECEx BVS 16.0005

PR 6241/..E only
20 and 21 II 1D Ex ta IIIC T160 °C Da Ex ta IIIC T160°C Da TÜV 03 ATEX 2301X IECEx TUN 17.0025X all PR 6241 without /..E
2 II 3G Ex nA IIC T6 Gc Manufacturer’s declaration all PR 6241 without /..E
22 II 3D Ex tc IIIC T85 °C Dc Manufacturer’s declaration all PR 6241 without /..E
FMus IS CL I, II, III, DIV 1, GP A, B, C, D, E, F, G Entity – 4012 101 5688

NI CL I, II, III, DIV 2, GP A, B, C, D, E, F, G NIFW – 4012 101 5688

T4A Ta= -30 °C to 70 °C; T5 Ta= -30 °C to 55 °C

FM17USO276 all PR 6241 without /..E
FMca IS CL I, II, III, DIV 1, GP A, B, C, D, E, F, G Entity – 4012 101 5688

NI CL I, II, III, DIV 2, GP A, B, C, D, E, F, G NIFW – 4012 101 5688

T4A Ta= -30 °C to 70 °C; T5 Ta= -30 °C to 55 °C

FM17CA0138 all PR 6241 without /..E

Ordering information

Compression S-type load cells OIML R60, accuracy class D1 = 0.04%
Type Maximum capacity Order number
PR 6241/12D1 100 kg 940526241121
PR 6241/22D1 200 kg 940526241221
PR 6241/32D1 300 kg 940526241321
PR 6241/52D1 500 kg 940526241521
PR 6241/13D1 1 t 940526241131
PR 6241/23D1 2 t 940526241231
PR 6241/33D1 3 t 940526241331
PR 6241/53D1 5 t 940526241531
Compression S-type load cells OIML R60, accuracy class D1 = 0.04%, ex-version
Type Maximum capacity Order number
PR 6241/12D1E 100 kg 940566241121
PR 6241/22D1E 200 kg 940566241221
PR 6241/32D1E 300 kg 940566241321
PR 6241/52D1E 500 kg 940566241521
PR 6241/13D1E 1 t 940566241131
PR 6241/23D1E 2 t 940566241231
PR 6241/33D1E 3 t 940566241331
PR 6241/53D1E 5 t 940566241531
Precision compression S-type load cells OIML R60, accuracy class C3 = 0.015%
Type Maximum capacity Order number
PR 6241/22C3 200 kg 940526241227
PR 6241/32C3 300 kg 940526241327
PR 6241/52C3 500 kg 940526241527
PR 6241/13C3 1 t 940526241137
PR 6241/23C3 2 t 940526241327
PR 6241/33C3 3 t 940526241337
PR 6241/53C3 5 t 940526241537
Precision compression S-type load cells OIML R60, accuracy class C3 = 0.015%, ex-version
Type Maximum capacity Order number
PR 6241/22C3E 200 kg 940566241227
PR 6241/32C3E 300 kg 940566241327
PR 6241/52C3E 500 kg 940566241527
PR 6241/13C3E 1 t 940566241137
PR 6241/23C3E 2 t 940566241327
PR 6241/33C3E 3 t 940566241337
PR 6241/53C3E 5 t 940566241537
Precision compression S-type load cells OIML R60, accuracy class C6 = 0.008%
Type Maximum capacity Order number
PR 6241/22C6 200 kg 940526241228
PR 6241/32C6 300 kg 940526241328
PR 6241/52C6 500 kg 940526241528
PR 6241/13C6 1 t 940526241138
PR 6241/23C6 2 t 940526241238
PR 6241/33C6 3 t 940526241338
Precision compression S-type load cells OIML R60, accuracy class C6 = 0.008%, ex-version
Type Maximum capacity Order number
PR 6241/22C6E 200 kg 940566241228
PR 6241/32C6E 300 kg 940566241328
PR 6241/52C6E 500 kg 940566241528
PR 6241/13C6E 1 t 940566241138
PR 6241/23C6E 2 t 940566241238
PR 6241/33C6E 3 t 940566241338
Precision compression S-type load cells OIML R60, converter Connexx®
Type Description Order number
PR 6241/DIGITAL KIT The converter Connexx® can only be ordered with the above-mentioned load cells from the PR 6241 series. When ordering, both order numbers need to be given. 940514100000

Additional information:
The converters Connexx® can be used with the Indicator X3.
To do this, the indicator X3 must be equipped with a CANopen interface card PR 5510/05 (940535510051).

Precision compression S-type load cells OIML R60, converter Connexx® – application sets
Type Description Order number
PR 6154/03 Application set for 3 load cells.

Includes: 2 × PR 6155/05, 1 × PR 6152/25, 1 × PR 6153/99

940536154031
PR 6154/04 Application set for 4 load cells.

Includes: 3 × PR 6155/05, 1 × PR 6152/25, 1 × PR 6153/99

940536154041
Precision compression S-type load cells OIML R60, converter Connexx® – load cell accessories
Type Description Order number
PR 6152/10 Connection cable between Connexx® and X3 (10 m) 940536152101
PR 6152/25 Connection cable between Connexx® and X3 (25 m) 940536152251
PR 6152/40 Connection cable between Connexx® and X3 (40 m) 940536152401
PR 6153/98 Divided cable gland 940536153981
PR 6153/99 Terminal resistor for Connexx®, M12 940536153991
PR 6155/05 Connection cable between Connexx® and Connexx® (5 m) 940536155051
PR 6155/10 Connection cable between Connexx® and Connexx® (10 m) 940536155101
 

Mounting kits PR 6041 without constrainer

Type Order number
PR 6041/04N 940536041041
PR 6041/30N 940536041301
PR 6041/30S 940536041302
PR 6041/40N 940536041401
PR 6041/40S 940536041402
PR 6041/51S 940536041512
Mounting kits PR 6043 –

MiniFLEXLOCK with integrated constrainer

Type Order number
PR 6043/30N 940536043301
PR 6043/30S 940536043302
PR 6043/40N 940536043401
PR 6043/40S 940536043402
Mounting kits PR 6044 – MaxiFLEXLOCK with additional functions
Type Order number
PR 6044/00S 940536044002
PR 6044/10S 940536044102
PR 6044/01S 940536044012
PR 6044/11S 940536044112
 

Mounting kits PR 6043 – load discs

Type Order number
PR 6043/31S 940536043312

 

Reviews

There are no reviews yet.

Be the first to review “Minebea intec / Load cell PR 6241, converter Connexx® and mounting kits PR 6041/PR 6043/PR 6044 For the precise weighing of process vessels”