FEATURES

  • Designed for Metrology Lab Use
  • Meets ASTM E74 Class A to 2% of Range, Class AA to 10% of Range (When Calibrated with Primary Standards)
  • Lowest Creep
  • Lowest Reproducibility Errors
  • Includes ASTM E74 Transfer Standard Tension & Compression Calibration
  • Temperature Compensated
  • Proprietary Modulus Compensated Strain Gages
  • Low Sensitivity to Rotation Error

SPECIFICATIONS

Model

Capacity      U.S. lbf.

CS10 300, 500,

b       1K, 2K

CS10

5K, 10K

CS20

25K, 50K

CS22

60K

CS30

100K

PARAMETERS                Metric

kN

1.5, 2.5, 5, 10

25, 50

100, 250

270

300, 500

ACCURACY

 

 

 

 

 

Lower Load Limit, % Rated Output

2.000

2.000

2.000

2.000

2.000

Static Error Band, % Rated Output

+/- 0.020

+/- 0.025

+/- 0.025

+/- 0.030

+/- 0.040

Nonlinearity, % Rated Output

+/- 0.020

 

+/- 0.030

+/- 0.030

+/- 0.030

+/- 0.045

Hysteresis, % Rated Output

+/- 0.020

+/- 0.040

+/- 0.045

+/- 0.040

+/- 0.045

Nonrepeatability, % Rated Output

+/- 0.005

+/- 0.005

+/- 0.005

+/- 0.005

+/- 0.005

Creep, % in 20 minutes

+/- 0.010

+/- 0.010

+/- 0.010

+/- 0.010

+/- 0.010

Off-Center Load Sensitivity, % / inch

+/- 0.050

+/- 0.050

+/- 0.050

+/- 0.050

+/- 0.050

Side Load Sensitivity, %

+/- 0.050

 

+/- 0.050

+/- 0.050

+/- 0.050

+/- 0.050

Zero Balance, % Rated Output

+/- 1.0

+/- 1.0

+/- 1.0

+/- 1.0

+/- 1.0

TEMPERATURE

 

 

 

 

 

Range, Compensated, °F

+15 to +115

+15 to +115

+15 to +115

+15 to +115

+15 to +115

Range, Compensated, °C

-10 to +45

-10 to +45

-10 to +45

-10 to +45

-10 to +45

Range, Operating, °F

-65 to +200

-65 to +200

-65 to +200

-65 to +200

-65 to +200

Range, Operating, °C

-55 to +90

-55 to +90

-55 to +90

-55 to +90

-55 to +90

Effect on Sensitivity, % Rdg / °F

+/- 0.0008

+/- 0.0008

+/- 0.0008

+/- 0.0008

+/- 0.0008

Effect on Sensitivity, % Rdg / °C

+/- 0.0015

+/- 0.0015

+/- 0.0015

+/- 0.0015

+/- 0.0015

Effect on Zero, % Rated Output / ° F

+/- 0.0008

+/- 0.0008

+/- 0.0008

+/- 0.0008

+/- 0.0008

Effect on Zero, % Rated Output / ° C

+/- 0.0015

+/- 0.0015

+/- 0.0015

+/- 0.0015

+/- 0.0015

ELECTRICAL

 

 

 

 

 

Rated Output, mV/V, Nominal

2

4

4

4

4

Excitation Voltage, VDC Maximum

20

20

20

20

20

Input Resistance, Ohms

350 +40 / -3.5

350 +40 / -3.5

350 +40 / -3.5

350 +40 / -3.5

350 +40 / -3.5

Output Resistance, Ohms

350 +/- 3.5

350 +/- 3.5

350 +/- 3.5

350 +/- 3.5

350 +/- 3.5

Insulation Resistance, Megohms

5000 @ 50 VDC

5000 @ 50 VDC

5000 @ 50 VDC

5000 @ 50 VDC

5000 @ 50 VDC

Bayonet Connector

PT02E-10-6P (MS3112E-10-6P)

MECHANICAL

 

 

 

 

 

Safe Overload Range, % Rated Capacity

+/- 150

+/- 150

+/- 150

+/- 150

+/- 150

Weight, lb

2.8

6.9

23.6

29.6

57.8

Weight, kg

1.3

3.1

10.8

13.4

26.2

Flexure Material

Aluminum

Steel

Steel

Steel

Steel

 

DIMENSIONS

Model

CS10

CS10M

CS10

CS10M

CS20

CS20M

CS22

CS22M

CS30

CS30M

Capacity

300, 500

1K, 2K lbf

1.5, 2.5,

5, 10 kN

5K, 10K lbf

25, 50 kN

25K, 50K lbf

100, 250 kN

60K lbf

270 kN

100K lbf

300, 500 kN

Dimensions

inch

mm

inch

mm

inch

mm

inch

mm

inch

mm

1.

4.13

104.9

4.13

106.7

6.32

160.5

7.68

195

7.85

199.4

2.

R 2.53

R 64.3

R 2.53

R 64.3

R 3.50

R 88.9

R 3.50

R 88.9

R 4.50

R 114.3

3.

4.12

104.8

4.12

104.8

6.06

154.0

6.06

154.0

8.00

203.2

4.

2.50

63.5

2.50

63.5

3.50

88.9

4.50

114.3

4.50

114.3

5.

0.19

4.7

0.19

4.7

0.30

7.5

0.31

7.9

0.40

10.2

6.

0.03

0.8

0.03

0.8

0.03

0.8

0.03

0.8

0.03

0.8

7.

1.25

31.8

1.25

31.8

2.25

57.2

3.00

76.2

3.00

76.2

8.

1/2 – 20

0.88 Deep

M12 x 1.25

22.4 Deep

5/8 – 18

0.88 Deep

M16 x 2

22.4 Deep

1 1/4 – 12

1.50 Deep

M33 x 2

38.1 Deep

1 1/2 – 12

2.0 Deep

M36 x 2

30.8 Deep

1 3/4 – 12

1.75 Deep

M42 x 2

44.5 Deep

9.

1/2 – 20

M12 x 1.25

5/8 – 18

M16 x 2

1 1/4 – 12

M33 x 2

1 1/2 – 12

M36 x 2

1 3/4 – 12

M42 x 2

10.

0.75

19.0

0.75

19.0

1.50

38.1

1.50

38.1

1.75

44.5

11.

R 4.00

R 101.6

R 4.00

R 101.6

R 6.00

R 152.4

R 6.00

R 152.4

R 12.00

R 304.8

OPTIONS

Additional bridges

Output adjustment to standard Compression overload protection Integral cable

Custom connectors Connector Protectors

ACCESSORIES

Mating connector Loading fixtures Interconnect cable Shunt calibration resistor Instrumentation

CS

Reviews

There are no reviews yet.

Be the first to review “Tovey CS SERIES LOAD CELLS”