설명
Model F1102 Hydraulic compression force transducer
The hydraulic force transducer model F1102 serves to easily capture and display forces in applications of limited mounting space. With its plastic piston, the transducer is especially suited to measure compression forces occurring during closing of welding tongs.
Hydraulic force measurement is a simple way to capture and display the forces occurring in various applications.
The force measurement is made using the principle of hydraulics: The force acting on a piston leads to a pressure increase which can be visualised on a connected display instrument. The scale of the display instrument can be defined in various units (e.g. N, kN, kg, t).
Leak-tightness warranty
The warranty on leak tightness of the hydraulic force measuring unit was extended to 5 years. A force transducer that starts to leak within this period will be repaired free of charge.
Specifications per VDI/VDE/DKD 2638
| Parameter | Value |
|---|---|
| Rated force (Fnom) | 0 … 100 N up to 0 … 36 kN |
| Nominal size (NS) | 6 |
| Display – Standard | Pressure gauge 213.40 (NS 63) |
| Display – Option | Digital pressure gauge DG-10 |
| Optional (on request) | Pressure sensor |
| Relative linearity error (dlin) – Standard | ≤ ±1.6 % Fnom (analogue display) |
| Relative linearity error (dlin) – Option | ≤ ±0.5 % Fnom (pressure sensor / digital pressure gauge) |
| Limit force (FL) | 100 % Fnom |
| Breaking force (FB) | > 130 % Fnom |
| Rated displacement (snom) | < 0.5 mm |
| Rated temperature range (BT, nom) | -25 … +50 °C |
| Ingress protection (EN/IEC 60529) | IP65 |
| Case material | Stainless steel |
| Piston material | Plastic (Delrin®) |
| Mounting type – Standard | Adapter L = 50 mm |
| Mounting type – Option | Capillary / Measuring hose (“separation without any losses”) |
| Fill fluid | Glycerine / water (70 % / 30 %) |
| Weight | 0.6 kg (with pressure gauge 213.40 NS 63) |
| 1 kg (with digital pressure gauge DG-10) |
1) For rated forces below 500 N, the relative linearity error is ±1.6 % Fnom for all connected measuring instruments.
Dimensions in mm

| Rated Force (N/kN) | System Pressure (bar) | 213.40 | DG-10 | Measuring Hose DN2 (max. L m) | Capillary (max. L m) |
|---|---|---|---|---|---|
| 100 N | 1.6 | ● | – | – | – |
| 160 N | 2.5 | ● | – | – | – |
| 250 N | 4 | ● | – | – | 1.0 |
| 400 N | 6 | ● | – | 0.5 | 1.0 |
| 600 N | 10 | ● | – | 1.0 | 2.0 |
| 1 kN | 16 | ● | – | 1.0 | 2.0 |
| 1.2 kN | 20 | – | ● | 1.5 | 2.0 |
| 1.6 kN | 25 | ● | – | 1.5 | 2.0 |
| 2.5 kN | 40 | ● | – | 1.5 | 2.0 |
| 3.2 kN | 50 | – | ● | 2.0 | 2.0 |
| 4 kN | 60 | ● | – | 2.0 | 2.0 |
| 6 kN | 100 | ● | ● | 2.0 | 2.0 |
| 10 kN | 160 | ● | ● | 2.0 | 4.0 |
| 16 kN | 250 | ● | ● | 3.2 | 4.0 |
| 20 kN | 315 | ● | – | 3.2 | 4.0 |
| 25 kN | 400 | ● | ● | 3.2 | 6.0 |
| 36 kN | 600 | ● | ● | 3.2 | 6.0 |
Other rated loads and versions on request
= possible selection
1) For a rated forces below 500 N, the relative linearity error is ±1.6 % Fnom for all connected measuring instruments.
2) Relative linearity error < ±1.0 % Fnom