AEP transducers / C10-AMP / Amplified Force transducers

Force transducer suitable for material testing machines and  test benches in general.
High mechanical resistance.
Completely made of stainless steel.
Completely LASER welded.
Degree of protection IP67.
Long term stability.

Category:

설명

Force transducers for the measurement of static and dynamic loads in COMPRESSION an TENSION

Nominal Load from 10 kN to 1 MN
Linearity and Hysteresis 0,10% and 0,20%
Standard Output

 

Dimensions[mm]

LOAD CODE
10 kN CC1014010KNI15
25 kN CC1014025KNI15
50 kN CC1014050KNI15
100 kN CC10140100KNI15
250 kN CC10165250KNI15
500 kN CC10165500KNI15
750 kN CC10200750KNI15
1000 kN CC102001MNI15

IMPORTANT NOTE: The contact surfaces with force transducer must have a hardness HRC ³ 40 and a roughness Ra ≤1.6

Technical Data

ITEM 10–25–50 kN
100 kN
250–500 kN 750–1000 kN
NOMINAL LOAD 10–25–50 kN
100 kN
250–500 kN 750–1000 kN
LINEARITY ≤ ±0.10% F.S. ≤ ±0.20% F.S.
AMPLIFIED OUTPUT
Sensitivity Tolerance
4–20 mA (3 wires), ±5 V, ±10 V
< ±0.1%
NOMINAL POWER SUPPLY ±5 V → 12–28 Vdc
4–20 mA & ±10 V → 18–28 Vdc
MAX. ABSORPTION
LOADING RESISTANCE
a) tension
b) current
RESPONSE FREQUENCY
30 mA

Min. 3 kΩ (tension)
0–470 Ω (current)
~2 kHz

TEMPERATURE EFFECT (10°C)
a) Zero
b) Sensitivity
≤ ±0.028% F.S.
≤ ±0.024% F.S.
REFERENCE  TEMPERATURE
TEMPERATURE NOMINAL RANGE
SERVICE TEMPERATURE RANGE
STORAGE TEMPERATURE RANGE
+23°C
–10 ~ +40°C
–10 ~ +70°C
–20 ~ +80°C
MECHANICAL LIMIT VALUES
a) Service load
b) Max permissible load
c) Breaking load
d) Max transverse load
e) Max dynamic load
DISPLACEMENT AT NOMINAL LOAD
120%
150%
>300%
100%
75%
~0.17 mm ~0.20 mm ~0.25 mm
WEIGHT ~4.1 kg ~5.5 kg ~8.3 kg
PROTECTION CLASS (EN 60529)
EXECUTION MATERIAL
IP67
Stainless Steel
CABLE OUTPUT 5 m
PVC 105°C shielded cable Ø5.3 mm
4 × tinned Ø0.14 mm² conductors

kg calibration on request.

Electrical Connections

Shield connected to the transducers body.

상품평

아직 상품평이 없습니다.

“AEP transducers / C10-AMP / Amplified Force transducers”의 첫 상품평을 남겨주세요