설명
Force transducers for the measurement of static and dynamic loads in COMPRESSION.
Nominal Load from 50 kg to 5 tons
Linearity and Hysteresis 0,20%
SMALL SIZE
Standard Output 2mV/V

| CODE | Loading head |
| CTICCM35 | For Ø35 force transducer |
| CTICCM42 | For Ø42 force transducer |
Dimensions[mm]

| NOMINAL LOAD | CODE |
| 50 kg | CCM50K3 |
| 100 kg | CCM100K3 |
| 250 kg | CCM250K3 |
| 500 kg | CCM500K3 |
| 1000 kg | CCM1T3 |
| 2500 kg | CCM2T53 |

| NOMINAL LOAD | CODE |
| 50 kg | CCM4250K3 |
| 100 kg | CCM42100K3 |
| 250 kg | CCM42250K3 |
| 500 kg | CCM42500K3 |
| 1000 kg | CCM421T3 |
| 2500 kg | CCM422T53 |
| 3500 kg | CCM423T53 |
| 5000 kg | CCM425T3 |
Technical Data
| NOMINAL LOAD | 50 – 100 – 250 – 500 kg 1000 – 2500 – 3500 – 5000 kg |
|
| NOMINAL SENSITIVITY CALIBRATION TOLERANCE |
2 mV/V £ ± 0.2 % F.S. |
|
| TEMPERATURE EFFECT (10 °C): a) on zero b) on sensitivity |
£ ± 0.030 % F.S. £ ± 0.025 % F.S. |
|
| LINEARITY HYSTERESIS REPEATABILITY |
£ ± 0.20 % F.S. £ ± 0.20 % F.S. £ ± 0.05 % F.S. |
|
| INPUT RESISTANCE OUTPUT RESISTANCE I NSULATION RESISTANCE ZERO BALANCE RECOMMENDED SUPPLY VOLTAGE NOMINAL SUPPLY VOLTAGE MAXIMUM SUPPLY VOLTAGE |
400 ± 20 W 350 ± 2 W > 5 GW £ ± 1 % F.S. 10 V 1 – 15 V 18 V |
|
| MECHANICAL LIMIT VALUES: a) service load b) max permissible load c) breaking load d) max transverse load e) max permissible dynamic load DISPLACEMENT AT NOMINAL LOAD |
120 % F.S. 150 % F.S. >300 % F.S. 50 % F.S. 50 % F.S. ~ 0.2 mm |
|
| REFERENCE TEMPERATURE TEMPERATURE NOMINAL RANGE SERVICE TEMPERATURE STORAGE TEMPERATURE |
+23 °C -10 / +40 °C -20 / +70 °C -20 / +80 °C |
|
| WEIGHT PROTECTION CLASS (EN 60529) EXECUTION MATERIAL CABLE LENGTH |
~ 0.13 kg | ~ 0.20 kg |
| IP67 Acciaio inox / Stainless Steel 3 m |
||
Electrical Connections

PVC 70°C shielded cable, Ø 4.2mm with 4 tinned Ø 0.14mm2 conductors. Shield connected to the body of the transducer.


상품평