HISP4 Single Point Load Cell
Applications
• Multi-head fillers
• Bench Scales/Floor Scales
• Batching/Blending
• Filling/Dispensing/Dosing
The HISP4 comes in fthree different capacity models ranging from 20kg to 100kg (44 to 220 lbs). The HISP4 includes NTEP approval and is rated to Class III accuracy level, making it suitable for use in trade-approved applications. An electropolished surface finish provides an added benefit for clean operations such as is found in the food and drug industries.
IDEAL FOR FILLER APPLICATIONS
The HISP4 has threaded mounting bolt holes on both the top and the front making it ideal to design into multi-head fillers. It is also designed for incorporation into bench scales, platform scales and floor scales.
C2 CALIBRATION
HISP4 load cells are designed with Hardy’s exclusive C2® Calibration for automatic system calibration to save hours in commissioning and routine maintenance. By reading the calibration with certificate data embedded electronically in the C2 sensor, the load cell can be calibrated the touch of a button, avoiding a time consuming traditional calibration procedure. This is essential for multi-head fillers where access to the load cell for traditional calibration requires taking the machine apart.
The HISP4 has threaded mounting bolt holes on both the top and the front making it ideal to design into multi-head fillers.
Specifications
| SPECIFICATIONS | Units | HISP4 |
| Maximum Capacity (Emax) | kgs | 20/ 50/ 100 |
| Maximum Capacity (Emax) | lbs | 44/110/220 |
| Accuracy Class | Class III | |
| Max # verification intervals | nmax | 3000 |
| Min load cell verification interval | vmin | Emax / 11000 |
| Combined Error | %RO | ± 0.0200 |
| Non-Linearity | %RO | ± 0.0166 |
| Hysteresis | %RO | ± 0.0166 |
| Creep error (30 Minutes) / DR | %RO | ± 0.0166 |
| Temp effect on min dead load | %RO/10°C | ± 0.0140 |
| Temp effect on sensitivity | %RO/10°C | ± 0.0100 |
| Rated Output (RO) | mV/V | 2 ± 0.2% |
| Zero Balance | %RO | ± 5 |
| Excitation Voltage | V | 5-15 |
| Input Resistance | Ω | 1100 ± 50 |
| Output Resistance | Ω | 960 ± 50 |
| Insulation resistance (100VDC) | MΩ | ≥ 5000 |
| Load Cell Safe Load Limit | %Emax | 200 |
| Load Cell Ultimate Load Limit | %Emax | 300 |
| Load Cell Safe Side Load | %Emax | 100 |
| Maximum Platform Size | mm | 500 x 400 |
| Compensated Temp Range | °C | -10 …±40 |
| Operating Temp Range | °C | -40…±80 |
| Load Cell Materil | Stainless Steel 17-4PH (1.4548) | |
| Sealing | Complete Hermetic Sealing – Glass to Metal Header | |
| Cable Length | ft/m | 10 ft/3 m |
| Hazardous Certification | IS Class 1,2,3 Div 1 | |
| Manufacturer’s Warranty | Two Years |

| MODEL NUMBER | CAPACITY LBS/KG |
| HISP4-44LB | 44 LB/20KG |
| HISP4-110LB | 110 LB/50KG |
| HISP4-220LB | 220 LB/50KG |
Cable Length: 10 ft. (3 Meters)
Dimensions: H 1.575 in (40 mm)
x 5.906 in (150 mm) x W .984 in
(25 mm)
Mounting Bolts: M6 x .5 Grade
8.8 or M6 x 1.0 Grade 8.8
Bolt Torque: 7 ft-lbs or 10 [Nm]
| C2 WIRE COLOR CODE FLAG LABEL IS FOUND APPROX. 10 IN. FROM END OF SENSOR’S CABLE | |
| EXCITATION + | RED |
| EXCITATION – | BLACK |
| SIGNAL + | GREEN |
| SIGNAL – | WHITE |
| C2+* | GRAY |
| C2 –* | VIOLET |
| SHIELD | YELLOW |
WARNING: NEVER cut load sensor cable